Abstract:
THE MEASUREMENT OF THE ISOTOPE RATIOS AND CONCENTRATIOS OF ZINC BY THERMAL IONIZATION MASS SPECTROMETRY USING DUOBLE ISOTOPE DILUTIONDENG ZHONGGUO(China Institute of Atomic Energy, Beijing, 102413)GUO DINGSHUN(Beijing Institute of Uranium Geology,100029)ABSTRACT The isotope ratios and concentrations of zinc are measured by silicagel-thermal ionizationmass spectrometry using the double isotope spikers. The double isotope spikers (~(70)Zn and ~(67)Zn-enriched isotopes) are used to correct the isotope mass fractionation for the zinc isotope ratios,andto certifv the zinc concentrations in the unknown samples. The zinc concentrations of these doubleisotope spikers are surveyed by a spiker made of pure(99. 99%)natural zinc metal powder. Thecorrecting factors (f_a、f_tandf_n)of the zinc isotope ratios in the spiked mixture,spike and un-spiked samples for the isotope mass fractionation,and the spike-to-unspiked ratios(X_r) of thezinc isotope r in the spiked mixture samples can be obtained to solve the matrix equations by nu-merical approximation.The natural zinc isotope ratios are:~(64)Zn/~(67)Zn=11. 8498,~(66)Zn/~(67)Zn=6.7977,~(68)Zn/~(67)Zn=4.5730 and~(70)Zn/~(67)Zn=0.1520. The uncertainies determined of the iso-tope ratios and concentrations of zinc are ±0. 16%and ±0.31%,respectivlely.Key words Double isotope dilution DOuble isotope spiker Isotope mass fractionation Matrix equation