Abstract:
THE IDENTIFICATION AND QUANTITATION FOR PERTECHNETATE IN THE ELUENT OF ~(99)TC~m GENERATOR BY HPLCLIU NING;ZHANG SHUYUAN;JIN JIANNAN;CHENG HENGLIU;ZHOU MAOLUN(Institute of Nuclear Science and Technology,Sichuan University,Chengdu 610064)ZHANG LIZHU;LI MAOLIAN(Nuclear Power Institute of China,Chengdu 610005)ABSTRACT The reversed phase HPLC is used to identify and quantify the pertechnetate in the eluent of~(99) Tc~m generator.The retention time of TcO is(3.00± 0.05)min and the detection limit forTcO is 1.5×10~(-8)mol/l.The Chromatorgraphic peak hights and concentration of TcO have agood linear relation with a relation coefficient of 0.9991.The HPLC procedure is also suitable forthe detection of pertechnetate in the radiopharmaceutical kits labeling with ~(99)Tc~m.Key words ~(99)Tc~m -generator Pertechnetate HPLC