    Calibration of 242Pu Tracer by Mass Spectrometry

    • 242Pu is commonly employed as tracer for determination of 239Pu in environmental samples. The 239Pu concentration in Pu(SO4)2•4H2O solution was determined by isotope dilution-mass spectrometry (ID-MS) using a high pure 242Pu spike. Then the 242Pu tracer in our laboratary was calibrated by ID-MS using the calibrated 239Pu solution as spike. By measuring the atom ratio of 239Pu to 242Pu twice by ID-MS, the concentration and abundance of the 242Pu tracer in laboratary were finally calibrated. This method was proved to be efficient with high precision. The process could be finished within two days with an uncertainty of 0.75%. The calibrated 242Pu tracer in laboratory could meet the demand of high-precision analysis.