HE Chang-shui, CHEN Xi-lin, YANG Hong-guang. Absolute Measurement of Micro-Calorimeter for Low Activity Tritium[J]. Journal of Nuclear and Radiochemistry, 2016, 38(1): 47-51. DOI: 10.7538/hhx.2016.38.01.0047
    Citation: HE Chang-shui, CHEN Xi-lin, YANG Hong-guang. Absolute Measurement of Micro-Calorimeter for Low Activity Tritium[J]. Journal of Nuclear and Radiochemistry, 2016, 38(1): 47-51. DOI: 10.7538/hhx.2016.38.01.0047

    Absolute Measurement of Micro-Calorimeter for Low Activity Tritium

    • For the absolute measurement requirement of trace tritium activity in ITER-TBS system, a micro-calorimeter was designed. The solid tritiated samples were used to demarcate the output thermal potential(Eout) and the input thermal power(P). The results show a good linear relationship between Eout and P, and the sensitivity coefficient reaching 0.13 V/W. The minimum detection limit of micro thermal meter is below 0.2 μW, and the calorimetric cup volume is more than 500 mL. The performance is expected to have a good application prospect for quantitative measurement of the trace tritium activity under various complicated situation in ITER-TBS system.