WEN Wen, WANG Meng, LOU Hai-lin, WEN Fu-ping. Measurement of 237Np in Urine Based on Alpha Spectrometer[J]. Journal of Nuclear and Radiochemistry, 2022, 44(5): 558-564. DOI: 10.7538/hhx.2022.YX.2021071
    Citation: WEN Wen, WANG Meng, LOU Hai-lin, WEN Fu-ping. Measurement of 237Np in Urine Based on Alpha Spectrometer[J]. Journal of Nuclear and Radiochemistry, 2022, 44(5): 558-564. DOI: 10.7538/hhx.2022.YX.2021071

    Measurement of 237Np in Urine Based on Alpha Spectrometer

    • 237Np is a highly toxic nuclide, which can cause internal exposure hazards after inhalation. The analysis of 237Np based on urine samples is an important method for routine monitoring, as well as a key method for studying the dose level of 237Np  in the human body of professional workers. The pretreatment process of 237Np in large-volume urine samples, pre-concentration by co-precipitation, and the effect of TEVA resin separation and purification and electrodeposition on the chemical recovery of 237Np were systematically studied and discussed. The research results show that the chemical recovery of the experimental method can reach 88.5%, and the detection limit of the method is 2.55×10-5 Bq/L. This analysis method meets the requirements of routine monitoring and is suitable for monitoring and evaluating 237Np in large-volume urine samples.