MA Shi-shen, WEN Xi-m. Analysis of Carbon Monoxide and Methane in Radioactive Argon[J]. Journal of Nuclear and Radiochemistry, 2006, 28(2): 90-93.
    Citation: MA Shi-shen, WEN Xi-m. Analysis of Carbon Monoxide and Methane in Radioactive Argon[J]. Journal of Nuclear and Radiochemistry, 2006, 28(2): 90-93.

    Analysis of Carbon Monoxide and Methane in Radioactive Argon

    • A gas loop that is being constructed in the Chinese Experimental Fast Reactor (CEFR) is simulated in a full scale. A sampling system and a feasible method are developed to realize a discontinuous online analysis for carbon monoxide and methane in argon cover gas system. By using a thermal conductivity detector in stead of a conventional hydrogen flame ionization detector, the exhaust gas from the outlet is directly discharged to a fume hood, thus eliminating the risk of contamination of the analytical laboratory by radioactivity that would be released in the latter case. The experimental results show that the adopted analytical procedure can meet the requirement of CEFR for routine analysis of CO and CH4:its precision is better than 3%, its accuracy is better than ±5% as evidenced by analyzing a set of CO and CH4 references, and its lower limit of measurement is 5×10-6.