Study on the Sampler for Radioxenon Isotopes and its Sampling Efficiency Calibration[J]. Journal of Nuclear and Radiochemistry, 2005, 27(2): 113-113.
    Citation: Study on the Sampler for Radioxenon Isotopes and its Sampling Efficiency Calibration[J]. Journal of Nuclear and Radiochemistry, 2005, 27(2): 113-113.

    Study on the Sampler for Radioxenon Isotopes and its Sampling Efficiency Calibration

    • An experimental sampler for rare gases (ESRG) which can collect radioxenon isotopes from atmosphere under normal temperature and low temperature is installed and studied. Its sampling efficiency is calibrated with ICP-MS (Inductively Coupled Plasma-Mass Spectrum) through analyzing the ~129Xe in the off-gas of ESRG. The main experimental results show that the sampling efficiency to xenon isotopes in ambient air is up to 40% when the adsorbent is effectually regenerated.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return